  • 25mW - 25W Avg, 60W Peak - laiaribaline võimsus-sensor ( Bird Electronic )

    Tootekood:

    Bird 5018D

    Kirjeldus:

    The 5018D RF Power Sensor never requires field calibration, only requires calibration once per year and is traceable to National Institute of Standards and Technology (NIST). The WPS measures True Average Power, Peak Power, and Duty Cycle directly with exceptional accuracy and uses these precise measurements to calculate a wide range of other important factors, such as VSWR, Return Loss, Reflection Coefficient, Crest Factor, Average Burst Power, and CCDF.

    Included Accessories: User Manual, VPM3 - Virtual Power Meter Software and Manual, 10 ft. USB Cable.


    • Frequency Range: 150 MHz - 4.0 GHz
    • Power Range: 100 mW - 25 Watts Avg. 60 Watts Peak
    • Insertion VSWR: <1.05 from 0.35 to 2.5 GHz; <1.10 from 2.5 to 4 GHz
    • Insertion Loss: <0.05 dB from 0.35 to 1.0 GHz; <0.1 dB from 1 to 4 GHz
    • Directivity: 30 dB up to 3.0 GHz; 28 dB from 3.0 to 4.0 GHz
    • Average power:
    Average Forward Power Range: 100 mW - 25 Watts Avg, 60 Watts Peak *Accuracy, Average Forward Power: ± 4% of reading, + 0.008 W
    Minimum Forward Power for Reflected Measurement: 0.1W
    Return Loss: 0.0 to 23 dB
    VSWR: 1.15 to 99.9
    • Burst average power:
    Burst Average Forward Power Range: .7 W - 25 Watts Avg
    Burst Width: 1 μs to 5 ms
    Repetition Rate: 5 Hz, Min
    Duty Cycle (D): .002 to 1.0
    *Accuracy, Burst Average Power: ± 6% of reading, + 0.008 Watts
    Peak Envelope Power Range: 0.7 - 60 W
    • Accuracy---Peak Envelope Power:
    burst width > 200 μs: ± 7% of reading, + 0.05 W
    1 μs < burst width < 200 μs: ± 10% of reading, + 0.10 W
    0.5 μs < burst width < 1 μs: ± 15% of reading, + 0.10 W
    burst width < 0.5 μs: ± 20% of reading, + 0.10 W
    • Crest Factor:
    Crest Factor Measurement Range: 25 mW - 25 Watts
    *Accuracy, Crest Factor: Linear Sum of Peak and Average Power Accuracies
    • Complementary Cumulative Distribution Function:
    CCDF Measurement Range: 0.1 to 100%
    Threshold Measurement Range: 0.7 - 25 W
    Measurement Uncertainty: ± 2%
    *Level Set Accuracy: As Peak Envelope Power Accuracy + 2.0%
    • Connectors: N Female (Both)
    • Power Supply: USB Port: less than one low-power USB load
    • DC Input Connector: 7-18 VDC at less than 0.1A
    • Impedance Ω: 50 (nominal)
    • Interfaces:
    DPM Interface: DB9 proprietary interface
    PC Interface (1): RS-232, 9600 Baud, no parity, 8 data bits, 1 stop bit, DB9
    PC Interface (2): USB 2.0 Type B
    • Weight: 1.2 lbs., max.
    • Dimensions H x W x D: 4.8" x 4.6" x 1.3" (122 x 117 x 33 mm)
    • Data Logging: Requires 5000-XT or VPM3
    • more information: 5018